Introduction – Labconco Protector Multi-Hazard Glove Box 5065502 User Manual

Page 6

Advertising
background image

INTRODUCTION

7

Performance

The Labconco Protector Multi-Hazard Glove Box has been carefully designed, constructed

and tested in order to provide an effective physical barrier between the laboratory and the glove
box interior. This barrier and the filtering system, provides a clean air enclosure for protection of
specimens and materials. It also provides protection to the technician from contaminated materials
and specimens.

Environmental Conditions

This equipment is designed to be safe under the following conditions:

• Indoor use
• Altitude up to 6562 Ft. (2000m)
• Temperature 41° to 104° (5° to 40°)
• Maximum relative humidity 80% for temperatures up to 88°F (31°C) decreasing linearly to

50% relative humidity at 104° (40°C)

• Main supply voltage fluctuations not to exceed ± 10% of the nominal voltage
• Transient over voltages according to Installation Categories II (Over voltage Categories per

IEC 1010)

• Pollution degrees 2 (Normally only non-conductive foreign matter, solid, liquid, or gaseous

(ionized gases), that may produce a reduction of dielectric strength or surface resistivity
occurs. Occasionally, however, a temporary conductivity caused by condensation must be
expected), in accordance with IEC 664