Troubleshooting

Manikin Troubleshooting

• Manikin will not connect:

o Check that the manikin is completely

turned off before searching. The
program will turn it on.

o Check cable or if you have extras, try to

change the cable.

o Try to disconnect other devices or

programs using the serial / usb ports.

• Ventilation feedback doesn’t correspond to

evaluation:

o Check that the lungs are installed

correctly.

o Try to change the lungs.
o Check that the face is connected

properly to the lung tubes.

• Need to hyperextend the neck when performing

ventilations:

o Check the technique by consulting the

introduction videos.

o Check the airway tubes in the neck.