11 Re-assembling the Resuscitator

To re-assemble the resuscitator, perform the steps described under Section 9 - Dis-assembling
the Resuscitator, in reverse.

12 Storage

For clinical use, store according to local protocol. For training, keep in the supplied
resuscitator’s storage pouch.

13 Glossary of symbols

0434 This product is in compliance with the essential requirements of Council

Directive 93/42/EEC as amended by Council Directive 2007/47/EC.

LATEX

This product does not contain latex.

14 Technical Information

Product meets ISO 10651-4:2002/EN ISO 10651-4:2009, Lung ventilators- Particular
requirements for operator-powered resuscitators
Operating Temperature:

- 18 °C to 50 °C

Storage Temperature:

- 40 °C to 60 °C

Tidal volume: 161 ml* +/- 15 ml (standard deviation) at room temperature
* In sub-zero temperatures, the tidal volume may be approx. 20% less.
Expiratory Resistance:

<2.5 cm H

2

O at 5 LPM

Inspiratory Resistance:

<0.5 cm H

2

O at 5 LPM

Dead Space:

4 ml (water volume)

Materials

Hard plastic components: Polysulfone (PSU)
Soft plastic components: Silicone rubber (SI)
Spring: Stainless Steel

External dimensions (with mask): Approx. 220 mm x 70 mm x 120 mm
Mass (with mask size 1): Approx 170 grams