8 Using the Resuscitator

Note: Safety valve: The resuscitator has a pressure release (pop-off) valve which releases air

when pressure to the patient exceeds 30-40 cmH

2

O. The valve opens to reduce the risk of

stomach distention and barotrauma. A hissing sound can be heard when the valve opens.

This safety valve may be overridden if more pressure to the patient is needed.
To override: Press downwards on the pressure release valve.

Note: For ventilation training with the NeoNatalie Newborn Simulator, use the largest mask

(no.1). For ventilation of a real patient, use the mask size that provides the best seal to the

patient’s face.

Note: If the Patient Valve becomes contaminated with vomitus shake free any contaminant

and squeeze the silicone bag to deliver several sharp breaths through the Patient Valve to

expel the contaminant. - If contaminant does not clear; disassemble the Patient Valve and

rinse.

Note: If any components are loose, tighten or reassemble the device and test in accordance

with section 7.

1. Slightly extend the

patient’s head and neck.

2. Cover the patient’s

mouth and nose
completely with the
mask. Press downwards
to create a seal between
the mask and the
patient’s face.

3. Squeeze the bag until

chestrise is observed.
Repeat this procedure at
the recommended rate.