5

6 General Hygiene

During Training
The resuscitator must be kept clean at all times:
• Wipe all surfaces clean with a wet cloth before and after each training session.

• If necessary use warm, soapy water. Wash hands between users.

For Patient Use
After each use, dis-assemble the resuscitator:
• Wash and scrub all parts in soapy water.

• Rinse in clean water.

• Sterilize or disinfect by using one of the methods as described in the section 10 -

Cleaning After Use.

Note: This product must be cleaned before first patient use. See section 10 - Cleaning After Use.

7 Check Before Each Use

To ensure proper functioning of the resuscitator, check the following functions before each use:

1. To ensure that air is delivered to the patient
Squeeze the bag and ensure the yellow lip valve opens with each squeeze. This indicates that
air can be successfully delivered to a patient.

Check that the

bag re-inflates

Valve opens

Squeeze the bag. Keeping the bag squeezed, seal the patient port connector against the
palm of one hand.

Release the bag, checking that it re-inflates quickly after release.