Troubleshooting – Laerdal SimJunior Standard User Manual

Page 26

Advertising
background image

24

Troubleshooting

• Check that there are no obstructions inside the manikin airways

which may block air flow.

• If there is no change when adjusting lung compliance, contact

Laerdal Technical Service.

• If there is no change in lung resistance, contact Laerdal Technical

Service.

Mechanical Noise during Auscultation

In the Instructor Application, click <Auscultation focus>.

Pulses (Radial and Brachial)

Problem

• Cannot feel pulses

Possible Solutions

• Skin may be too tight over pulse units – re-adjust skin and

reboot.

• Ensure blood pressure is set to at least 60 systolic.

• Ensure compressor is ON.

Note: If the radial pulse is set below 60, peripheral pulses

may disappear.

Manikin Shutdown

Problem

• The manikin is unresponsive.

Solution

1. Power off the Link Box.

2. Turn the SimPad off.

3. Exit any SimJunior software on the optional Instructor PC and

the Patient Monitor.

4. Then restart everything.

If the manikin is still unresponsive, contact your Laerdal representative.