Laerdal SimMan User Manual

Page 5

Advertising
background image

E

N

G

L

IS

H

5

12) Attach the thigh electrical connector to the open mating

connector in the pelvis. Right leg wire goes to the wire on the
right side, and left goes to left.

13) Repeat with the other leg.
14) Gently replace the belly.

To remove leg(s), reverse the procedure.

Link Box
The Link Box connects the manikin to the computer.
1) Attach manikin cable to right lower side of manikin’s torso and

to the connector marked “Manikin” on the back of the Link Box.

2) Connect serial cable to connector marked “PC” on the back of

the Link Box and to serial port on back of your computer.

3) Connect clear tubing from blood pressure cuff to the inlet

marked “BP cuff” on the back of the Link Box.

4) Connect the SpO

2

cable to the SpO

2

connector on the back

of the Link Box.

5) The audio cable has three ends. Connect the RCA / phono plug

to the connector marked "Audio input" on the back of the link
box. The stereo jack with single wire shall be connected to the
Simulated Patient Monitor, and the stereo mini jack with double
wire shall be connected to the computer's headphone output.
For illustrations please see the accompanying Install Guide.
If a wireless microphone is used, the receiver should be
connected directly to the Link Box connector marked "Audio
input" instead of the audio cable described above (pre-defined
vocal sounds will not be possible to use in parallel)

6) Plug the Link Box AC power cable into a power supply

(110-240 V AC). If you are using the Portability Kit, attach to
the 12 V DC input according to the Portability Kit instructions.

7) Connect external speakers, if used, to connector marked

"Ext. speaker" on the back of Link Box.

USB Hub
A USB hub is supplied with the system to provide additional USB
ports. Connect the USB Hub to one of the USB connectors at the
back of the PC. Attach the AC adapter to the USB Hub and to a
wall outlet (110 or 230-240V AC).

SpO2 Probe
The SpO

2

finger probe provides enhanced realism for the care

provider by simulating the use of an actual pulse oximetry probe.
When the probe cable is connected to the Link Box, SpO

2

will not

be displayed until the probe is placed on one of the manikin’s
fingers (means that “light” is blocked). If the probe is not attached
to the Link Box, SpO

2

will only be displayed automatically when

selected by the operator through the instructor panel on the PC.

Blood system
The blood system includes a blood bag with tubing that can be
connected to the veins and be altered by varying the elevation of
the blood bag.

Air and CO

2

Source (optional)

Option 1: Compressor Unit

If you have purchased Laerdal SimMan with a Compressor Unit,
attach one end of the double lumen tube into the compressor and
the other to the right axillary side of the manikin.

Start procedure:
1) Check that Power switch (4) is set to off position “0”.
2) Plug power supply cable into plug (5) in Compressor panel.
3) Plug power supply cable into power source.
4) Connect CO

2

supply to the “Input CO

2

” connector using

the black hosing attached (optional feature).

5) Check that Drain valve (1) is closed.
6) Close Air valve (2) and CO

2

valve (3)

7) Push Power switch (4) to on position “1”, the Compressor will

start and run for approx. 45 seconds to build up pressure in tank.

8) When Compressor stops, you can start to use the manikin.

Open the Air valve (2) and the CO

2

valve (3).

9) The Compressor will start and stop with different intervals

depending on consumption of air.

Stop procedure:
1) Push Power switch (4) to off position “0”, the Compressor will

stop if it is running.

2) Close Air valve (2) and CO

2

valve (3).

3) Open Drain valve (1) and drain air out of system.

Warning:
Do not open Compressor box when energized, dangerous voltage inside.
Only to be opened by authorized personnel.

Important:
If the Compressor Unit is stopped manually with the on/off switch, the
air tank must be drained before restart.

Advertising