LW Measurements WB 10000 User Manual

Page 13

Advertising
background image

. Output: 12V DC 150mA

!

Allowable ambient conditions

!

Temperature: 5°C - 40°C

Relative humidity: 25% - 85%, non-condensing

!

If you have any questions on the technical data or require detailed technical information on your balance,

please contact your technical representative.

!!

2.3.2 RS232 Data interface

!

1.

Connector: DB9

!

SCALE

PC

2

2

3

3

5

5

!!

2.

Transmissions Settings

!

Mode: Simplex Asynchronous Serial

Baud Rate: 9600

Data Bit: 8

Parity Bit: None

Stop Bit: 1

Data Format: ASCII

!

3.

Transmission Information Format: 20 Byte ,blank=20H

! !

W:

Start of Data Transmission=57H+3AH

P :

Polarity ‘+’ = 2BH = Positive

‘-’ = 2DH =Negative

DATA: ‘ 12.345’= 20H+20H+20H+20H+31H+32H+2EH+33H+34H+35H

UNIT: ‘lb’

=6CH+62H+20H+20H+20H

‘kg’

=6BH+67H+20H+20H+20H

CR:

= 0DH

LF:

= 0AH

!

1~2

3

4~13

14~18

19

20

‘W: ’

P

DATA

UNIT

CR

LF

!