LW Measurements WB 10000 User Manual

Page 20

Advertising
background image

!

4.2.2.3 Setting power-on mode

!

Press and hold HOLD key, then press ON/OFF key to turn on the Scale, display will show “A ON” or “A

OFF”, Press UNIT key three times, display will show “J ON” or “J OFF”, press ZERO key to select “J

ON” (means showing zero when scale is turned on) or “J OFF” (means showing the weight placed when

the scale is turned on).

!

!

!

!

4.2.3 Working mode settings

!!

- Press & hold UNIT and HOLD keys, then press ON/OFF key to turn on the scale

!

- The display will show Ani. Su, press UNIT key to enter, press ZERO key to select ON (means animal
weighing mode) or OFF (means normal weighing mode), press HOLD key to confirm.

!

- Press HOLD key again, display will show Ani. Kd, press UNIT key to enter, press ZERO key to select
anti-shock range between 50, 100, 200 and 400, press HOLD key to confirm.
*** This is the anti-shock range 50d, 100d, 200d or 400d. The smaller division range is suitable for smaller
animals, the larger division range for larger animals

!

- Press HOLD key again, display will show Ani. Fd, press UNIT key to enter, press ZERO key to select
filter range between 0.5, 3, 5 and 10, press HOLD key to confirm.
*** This is the filter range, 0.5d, 3d, 5d or 10d that the scale displays, for example, if we use LVS 700lb x
0.2lb for weighing a 300lb cow, if we set it to be 0.5d, then the scale will show 299.8lb, 300lb, 300.2lb..., if
we set it to be 3d, then the scale will show 299.4lb, 300lb, 300.6lb... 0.5d can get much accurate
weighing, but it won’t be as stable; 10d will not obtain such an accurate weighing result, but it will be
much more stable.

!

- Press HOLD key again, the display will show Ani. Ft, press UNIT key to enter, press ZERO key to select
filter time between 1, 2, 3 and 4, press HOLD key to confirm.
*** This is the filter time, 1s, 2s, 3s or 4s that will take to display the weighing. Weighing will be more
accurate if the user chooses 4 second (longer displaying time), but not so accurate if 1 second (faster
displaying time) is chosen.

!

Press ON/OFF key to turn the scale off.

!!

!

4.2.4 Hold function

!!

- Press ON/OFF key to turn on the scale, place the item on the platform.

!

- Wait for the readings stable, press HOLD key, the “

.

“ indicator will show

!

- Remove the item from the platform, the readings will still remain on the display

!

!