Adjusting setpoint <2 – Nor-Cal CDG 045 OP Lit User Manual
Page 29

CDG045-OP-LIT (2010-1)
29
For changing the threshold value, push the <ZERO> button
and keep it depressed. The threshold keeps changing from
the current value (ramp) until the button is released or until
the limit of the setting range is reached.
Keep the button depressed
max. ø1.1 mm
Push the <ZERO> button again:
Fine adjustment
within 0...3 s:
the zero adjustment value changes
by one unit
Change of direction
within 3...5 s:
the zero adjustment changes its
direction (the blinking frequency of
the <STATUS> LED changes
briefly)
If the <ZERO> button is released for more than 5 s, the
gauge returns the measurement mode.
The upper threshold is automatically set 1% F.S. above
the lower one (hysteresis).
Adjusting Setpoint <2>
Push the <SP> button twice (the LED <2> blinks). The adjust-
ment procedure is the same as for setpoint <1>.