2 zeroing the gauge – Nor-Cal CDG 100 OP Lit User Manual

Page 23

Advertising
background image

CDG100-OP-LIT (2010-1)

23

4.2

Zeroing the Gauge

The gauge is factory calibrated while "standing upright"
(

→ "Calibration Test Report").

We recommend performing a zero adjustment, when the

gauge is operated for the first time.

Due to long time operation or contamination, a zero drift could
occur and zero adjustment may become necessary.

For adjusting the zero, operate the gauge under the same con-
stant ambient conditions and in the same mounting orientation
as normally.

The output signal (measuring signal) is depending on the moun-
ting orientation. The signal difference between the vertical and
horizontal mounting orientation is:

F.S.

U / 90°

1000 Torr

≈2 mV

100 Torr

≈10 mV

10 Torr

≈50 mV

1 Torr

≈300 mV

0.1 Torr

≈1.8 V

If the gauge is operated via a controller, the zero of the

whole measuring system has to be adjusted on the
controller: first, adjust the zero of the gauge and then,
the zero of the controller.

Advertising