Measuring AC Amperage

CAUTION!

Do not attempt to make a current measurement with the
test leads. The 275 measures the current by clamping the
jaw around one conductor (wire). Clamping around more
than one wire will result in erroneous readings. Make sure
the temperature probe is NOT plugged in during this test.

Instrument set-up:

FUNC.

BLACK RED

MIN

MAX

TEST LEAD TEST LEAD

READING

READING

NOT USED NOT USED

0.01A

400A

Measurement Procedure:

1.

Disconnect power to circuit to be measured.

2.

Set rotary switch to

function.

3.

Clamp the jaws around one conductor of the circuit to
be measured. For best results, center the wire in the
jaw.

4.

Reconnect power to circuit to be measured.

5.

Read the current on the 275.

Optional Modes

(See Other Features Section)

D-H:

Freezes the reading on the LCD.

MAX/MIN:

Records minimum and maximum measure-

ments

P-H:

Displays peak amperage.

17

16

Measuring Resistance

WARNING!

Do not attempt to make resistance measurements with circuit
energized. For best results, remove the resistor completely from
circuit before attempting to measure it. Make sure the tempera-
ture probe is NOT plugged in during this test.

NOTE:

To make accurate low ohm measurements, short the ends of the
test leads together and press the REL button to store the reading.
This will deduct the stored value from subsequent measurements
eliminating the test lead resistance from the reading.

Instrument set-up:

FUNC.

BLACK RED

MIN

MAX

TEST LEAD TEST LEAD

READING

READING

COM

V/

0.1

40.00M

Measurement Procedure:

1.

Disconnect power to circuit to be measured.

2.

Plug black test lead into the COM input jack.

3.

Plug red test lead into V/

input jack.

4.

Set the rotary switch to the

function.

5.

Connect test leads to circuit to be measured.

6.

Read the resistance value on the 275.

Optional Modes

(See Other Features Section)

D-H:

Freezes the reading on the LCD.