Accuracy ±( [ % of Reading ] + [ Counts ] )

Function

Range

1

Resolution

Model 175

Model 177

Model 179

AC Volts

2, 3

600.0 mV

6.000 V

60.00 V

600.0 V

1000 V

0.1 mV

0.001 V

0.01 V

0.1 V

1 V

1.0 % + 3

(45 Hz to 500 Hz)

2.0 % + 3 (500 Hz to 1 kHz)

1.0 % + 3

(45 Hz to 500 Hz)

2.0 % + 3 (500 Hz to 1 kHz)

1.0 % + 3

(45 Hz to 500 Hz)

2.0 % + 3 (500 Hz to 1 kHz)

DC mV

600.0 mV

0.1 mV

0.15 % + 2

0.09 % + 2

0.09 % + 2

DC Volts

6.000 V

60.00 V

600.0 V

0.001 V

0.01 V

0.1 V

0.15 % + 2

0.09 % + 2

0.09 % + 2

1000 V

1 V

0.15 % + 2

0.15 % + 2

0.15 % + 2

Continuity 600

Ω

1

Ω

Meter beeps at < 25

Ω

, beeper turns off at > 250

Ω

; detects opens or shorts of 250

μ

s or

longer.

Ohms

600.0

Ω

6.000 k

Ω

60.00 k

Ω

600.0 k

Ω

6.000 M

Ω

50.00 M

Ω

0.1

Ω

0.001 k

Ω

0.01 k

Ω

0.1 k

Ω

0.001 M

Ω

0.01 M

Ω

0.9 % + 2

0.9 % + 1

0.9 % + 1

0.9 % + 1

0.9 % + 1

1.5 % + 3

0.9 % + 2

0.9 % + 1

0.9 % + 1

0.9 % + 1

0.9 % + 1

1.5 % + 3

0.9 % + 2

0.9 % + 1

0.9 % + 1

0.9 % + 1

0.9 % + 1

1.5 % + 3

Diode test

2.400 V

0.001 V

1 % + 2

Capacitance

1000 nF
10.00

μ

F

100.0

μ

F

9999

μ

F

4

1 nF

0.01

μ

F

0.1

μ

F

1

μ

F

1.2 % + 2

1.2 % + 2

1.2 % + 2

10 % typical

1.2 % + 2

1.2 % + 2

1.2 % + 2

10 % typical

1.2 % + 2

1.2 % + 2

1.2 % + 2

10 % typical

AC Amps

5

(True RMS)
(45 Hz to 1

kHz)

60.00 mA

400.0 mA

6

6.000 A

10.00 A

7

0.01 mA

0.1 mA

0.001 A

0.01 A

1.5 % + 3

1.5 % + 3

1.5 % + 3

1. All AC voltage and AC current ranges are specified from 5 % of range to 100 % of range.

2. Crest factor of

3 at full scale up to 500 V, decreasing linearly to crest factor

1.5 at 1000 V.

3. For non-sinusoidal waveforms, add -(2% reading + 2% full scale) typical, for crest factors up to 3.

4. In the 9999

μ

F range for measurements to 1000

μ

F, the measurement accuracy is 1.2 % + 2 for all models.

5. Amps input burden voltage (typical): 400 mA input 2 mV/mA, 10 A input 37 mV/A.

6. 400.0 mA accuracy specified up to 600 mA overload.

7. > 10A unspecified.