Fluke 179 User Manual

Page 18

Advertising
background image

Models 175, 177 & 179
Users Manual

14

Function Overload

Protection

1

Input Impedance

(Nominal)

Common Mode Rejection Ratio

(1 k

Ω

Unbalanced)

Normal Mode Rejection

Volts AC

1000 V RMS

> 10 M

Ω

< 100 pF

> 60 dB @ DC, 50 Hz or 60 Hz

Volts DC

1000 V RMS

> 10 M

Ω

< 100 pF

>120 dB @ DC, 50 Hz or 60 Hz

> 60 dB @ 50 Hz or 60 Hz

mV/T

1000 V RMS

2

> 10 M

Ω

< 100 pF

>120 dB @ DC, 50 Hz or 60 Hz

> 60 dB @ 50 Hz or 60 Hz

Full Scale Voltage To:

Open Circuit Test

Voltage

600 k

Ω

50 M

Ω

Short Circuit Current

Ohms/Capacitance

1000 V RMS

2

< 8.0 V DC

< 660 mV DC

< 4.6 V DC

< 1.1 mA

Continuity/Diode

test

1000 V RMS

2

< 8.0 V DC

2.4 V DC

< 1.1 mA

1. 10

7

V-Hz maximum.

2.

For circuits < 0.3 A short circuit. 660 V for high energy circuits.


Function Overload

Protection Overload

mA

Fused, 44/100 A, 1000 V FAST Fuse

600 mA overload for 2 minutes maximum, 10

minutes rest minimum

A

Fused, 11 A, 1000 V FAST Fuse

20 A overload for 30 seconds maximum, 10

minutes rest minimum