Fluke 179 User Manual

Page 17

Advertising
background image

True RMS Multimeters

Specifications

13

Accuracy ±( [ % of Reading ] + [ Counts ] )

Function

Range

1

Resolution

Model 175

Model 177

Model 179

DC Amps

4

60.00 mA

400.0 mA

6

6.000 A

10.00 A

7

0.01 mA

0.1 mA

0.001 A

0.01 A

1.0 % + 3

1.0 % + 3

1.0 % + 3

Hz
(AC- or DC-

coupled, V or

A

2, 3

input )

99.99 Hz

999.9 Hz

9.999 kHz

99.99 kHz

0.01 Hz

0.1 Hz

0.001 kHz

0.01 kHz

0.1 % + 1

0.1 % + 1

0.1 % + 1

Temperature

-40

°

C to +400

°

C

-40

°

F to +752

°

F

0.1

°

C

0.1

°

F

NA NA

1 % + 10

5

1 % + 18

5

MIN MAX AVG For DC functions, accuracy is the specified accuracy of the measurement function

±

12 counts for changes longer than 350 ms in

duration.
For AC functions, accuracy is the specified accuracy of the measurement function

±

40 counts for changes longer than 900 ms in

duration.

1. All AC voltage and AC current ranges are specified from 5 % of range to 100 % of range.
2. Frequency is specified from 2 Hz to 99.99 kHz in Volts and from 2 Hz to 30 kHz in Amps.
3. Below 2 Hz, the display shows zero Hz.
4. Amps input burden voltage (typical): 400 mA input 2 mV/A, 10 A input 37 mV/A.
5. Does not include error of the thermocouple probe.
6.

400.0 mA accuracy specified up to 600 mA overload.

7.

> 10A unspecified.