Fluke 189 User Manual

Page 94

Advertising
background image

Model 187 & 189
Users Manual

7-10

Function

Range

Resolution

Accuracy

Frequency

500.00 Hz

0.01 Hz

1

±

(0.0050 % + 1)

5.0000 kHz

0.0001 kHz

50.000 kHz

0.001 kHz

999.99 kHz

0.01 kHz

Duty Cycle

10.00% to 90.00 %

0.01 %

±

((voltage range/input voltage) X 300 counts)

5,6

Pulse Width

499.99 ms

0.01 ms

±

(3 % X (voltage range/input voltage) + 1 count)

5,6

999.9 ms

0.1 ms

Temperature

200 to +1350

°

C

0.1

°

C

±

( 1% of reading + 1

°

C)

2,3

328 to +2462

°

F

0.1

°

F

±

( 1% of reading + 1.8

°

F)

2,3

MIN MAX AVG

Response: 100 ms to 80 %

Specified accuracy

±

12 counts for changes > 200 ms

in duration. (

±

40 counts in AC for changes > 350 ms

and inputs > 25 % of range)

FAST MN MX

250

µ

s

4

Specified accuracy ±100 counts up to 5,000 count (full
range) reading. For higher peak readings (to 20,000
counts), specified accuracy

±

2% of reading.

1. Reading will be 0.00 for signals below 0.5 Hz.
2. Accuracy specification is relative to the user-adjustable temperature offset, and assumes ambient temperature stable to

±

1

°

C.

3. For ambient temperature changes of

±

5

°

C, rated accuracy applies after 1 hour.

4. For repetitive peaks; 2.5 ms for single events. Use DC function settings below 20 Hz. 50 mV range not specified.
5. Frequency greater than 5 Hz, except for VDC, 500 mVDC and 3000 mVDC functions; 0.5 Hz to 1 kHz. Signals centered around
trigger levels.
6. Range/input ratios also apply to current functions. 500 counts or 5 % for 10 A ranges.