Frequency counter sensitivity, Burden voltage (a, ma, µa) – Fluke 189 User Manual

Page 95

Advertising
background image

Specifications

Frequency Counter Sensitivity

7

7-11

Frequency Counter Sensitivity

Approximate VAC Sensitivity

(RMS Sine Wave)

1

Input Range

15 Hz to

100 kHz

2

500 kHz

2

VAC Bandwidth

3

Approximate VDC

Trigger Levels

1

VDC

Bandwidth

3

50 mV

5 mV

10 mV

1 MHz

-5 mV & 5 mV

1 MHz

500 mV

20 mV

20 mV

1 MHz

5 mV & 65 mV

1 MHz

3000 mV

500 mV

2000 mV

800 kHz

140 mV & 200 mV

90 kHz

5 V

0.5 V

2.0 V

950 kHz

1.4 V & 2.0 V

14 kHz

50 V

5 V

5.0 V

1 MHz

0.5 V & 6.5 V

> 400 kHz

500 V

20 V

20 V

1 MHz

5 V & 65 V

> 400 kHz

1000 V

100 V

100 V

> 400 kHz

5 V & 65 V

> 400 kHz

.

1. Maximum input = 10 x Range (1000 V max). Noise at low frequencies and amplitudes may affect accuracy.
2. Useable at reduced sensitivity to 0.5 Hz and 1000 kHz.

.

3. Typical frequency bandwidth with full scale (or maximum 2 X 10

7

V-Hz product) RMS sine wave.

Burden Voltage (A, mA, µA)

Function

Range

Burden Voltage (typical)

mA -

µ

A

500.00

µ

A

102

µ

V /

µ

A

5,000

µ

A

102

µ

V /

µ

A

50.000 mA

1.8 mV / mA

400.00 mA

1.8 mV / mA

A

5.0000 A

0.04 V / A

10.000 A

0.04 V / A