18.3.2 Performing a hot-air sterilization

Before carrying out the first hot-air sterilization, remove any protective lamination sheet from

the inner metal surfaces.

When starting a hot-air sterilization, all gas controls automatically become inactive.

The safety controller settings are inactive during sterilization. They become functional again

following abortion of the sterilization and/or restart of the unit at the main power switch.

•

Turn off the unit.

•

Empty the Permadry™ water pan.

WARNING

Danger of implosion.
Destruction of the incubator.

Empty the Permadry™ water pan before starting the hot-air sterilization.

•

Water pans and shelves must be inside the incubator.

•

Close the inner glass door and the outer unit door.

•

Turn on the unit.

•

Activate the sterilization procedure in the controller.



To start the hot-air sterilization cycle, go to

Menu > Hot-air sterilization

or

Quick menu > Hot-air sterilization

Fixed value

08.03.2013 05:05:06

Submenu “Hot-air sterilization”.
Select “Start”

and press the operating button.

..\ Hot-air sterilization

Start

Stop

Close