CB (E6) 09/2013

Page 48/144

6.5 Turning on / off the interior socket voltage (with optional interior socket)

For chambers equipped with the optional interior socket (chap. 15.4) you can turn on and off the voltage

of the interior socket.
To turn on or off the interior socket voltage, go to

Menu > Interior socket (optional)

Fixed value

08.03.2013 05:05:06

Submenu “Interior socket”.
Select the desired setting

and press the operating button
“Socket On” = socket with activated voltage

“Socket Off” = socket voltage-free

..\ Interior socket

Socket On

Socket Off

Close

Home

Go back to the initial view with “Home”.

This symbol on the controller display indicates that the interior socket is activated.


6.6 Performance during and after power failure

During a power failure, all controller functions are shut down. The gas inlet valves are closed so that no

gas can escape into the ambient air. The zero-voltage relay alarm output (9) (chap. 11.6) is switched to

alarm position for the whole duration of the power failure.
After the power returns, all functions return to the same status the chamber had before power failure. .

The controller continues to function in the original operating mode it was in previously before the power

failure occurred. In “fixed value” mode, the setpoints are immediately resumed. The power failure and

power return are noted in the event list.
If the CO

2

incubator had been running in sterilization mode, the sterilization process is cancelled and the

unit operates in “fixed value” operation mode with the setpoints already entered. All setpoint values

remain in memory.
If the temperature or concentration of CO

2

, and O

2

(unit with O

2

control) have dropped below the alarm

limits during power failure, confirm the alarms with the RESET button as soon as the correct values are

reached again (chap. 11.4).