3 aborting the hot-air sterilization prematurely, Caution – BINDER CB 60 User Manual

Page 113

Advertising
background image

CB (E6) 09/2013

Page 113/144


Opening the outer door during sterilization leads to aborting the automatic sterilization cycle.

CAUTION

Interruption of the temperature reaction time.
Ineffective sterilization.
∅

Do NOT open the unit doors during sterilization.

CAUTION

Glass door and inner chamber become hot during sterilization.
Danger of burning.
∅

Do NOT touch the glass door and inner surfaces during sterilization


18.3.3 Aborting the hot-air sterilization prematurely

Three events lead to aborting the automatic sterilization cycle prematurely:
•

A manual abortion via the controller menu

•

Opening the outer door

•

Turning off the incubator at the main power switch, or a power failure


To manually abort the hot-air sterilization cycle, go to

Menu > Hot-air sterilization

or

Quick menu > Hot-air sterilization

Sterilization

08.03.2013 05:05:06

Submenu “Hot-air sterilization”.
To abort the hot-air sterilization, select “Stop”

and press the operating button.

..\ Hot-air sterilization
Start

Stop

Close

Home

Sterilization

08.03.2013 05:05:06

Submenu “Stop hot-air sterilization”.
This is a security question.
To abort the hot-air sterilization, select “Really stop”

and press the operating button.

..\ Stop hot-air sterilization

Do not stop

Really stop

Close

Home