Error messages/troubleshooting – Beurer BC 19 User Manual

Page 22

Advertising
background image

22

• In order to retrieve measurement results again, press

the memory key M. Initially, the average value of the

last three blood pressure readings is displayed (Dis-

play ). Press the memory key M again to display the

latest measurements with date, time and graphic WHO

classification and to hear the values announced. You

can adjust the volume during the announcement by

pressing the volume key + and -.

• You can clear the memory by pressing and holding +

and - simultaneously for 5 seconds after pressing the

memory key M and selecting the user.

• If you wish to change the user memory, see “Select

memory”.

7. Error messages/troubleshooting

In case of faults, the _ message appears in the display.

Error messages may appear if:

• systolic or diastolic pressure could not be measured

(

or

appears on the display)

• systolic or diastolic pressure was outside the measure-

ment range ( or

Lo

appears on the display)

• the cuff is fastened too tightly or loosely (

Er3

or

Er4

appears on the display)

• the pump pressure is higher than 300 mmHg (

Er5

appears on the display)

• pumping up takes longer than 160 seconds (

Er6

appears on the display)

• there is a system or device error (

ErA

,

, Er7

or

Er8

appears on the display)

• the batteries are almost empty

.

In the above cases, you must repeat the measurement.

Make sure that the cuff tube is properly inserted and that

you do not move or talk.

Re-insert the batteries if necessary, or else replace

them.
Technical alarm – description
Should the recorded blood pressure (systolic or diasto-

lic) lie outside the limits specified in the section “Techni-

cal specifications”, the technical alarm will appear on the

display indicating either “ ” or “

Lo

”. In such cases, you

should seek medical assistance and check the accuracy

of your procedure.

The limit values for the technical alarm are factory set

and cannot be adjusted or deactivated. These alarm limit

values are accorded second priority under the standard

IEC 60601-1-8.

The technical alarm is a non-locking alarm and must not

be reset. The signal shown on the display will disappear

automatically after about 8 seconds.