Table 3
For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING

Guidance and manufacturer’s declaration – electromagnetic immunity
The BC 19 is intended for use in the electromagnetic environment specified below. The customer or the user of the BC 19 should

assure that it is used in such an environment.
Immunity test IEC 60601 test

level

Compliance

level

Electromagnetic environment – guidance

Radiated RF

IEC 61000-4-3

3 V/m 80 MHz

to 2.5 GHz

3 V/m

Portable and mobile RF communications equipment should be used no closer to

any part of the BC 19, including cables, than the recommended separation dis-

tance calculated from the equation applicable to the frequency of the transmitter.
Recommended separation distance:
d = 1.2

80 MHz to 800 MHz

d = 2.3

800 MHz to 2.5 GHz

Where P is the maximum output power rating of the transmitter in watts (W)

according to the transmitter manufacturer and d is the recommended separation

distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic

site survey,

a

should be less than the compliance level in each frequency range.

b

Interference may occur in the vicinity of equipment marked with the

following symbol:

NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.

NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection

from structures, objects and people.