Beurer BC 19 User Manual

Page 64

Advertising
background image

64

Table 2
For all ME EQUIPMENT and ME SYSTEMS

Guidance and manufacturer’s declaration – electromagnetic immunity
The BC 19 is intended for use in the electromagnetic environment specified below. The customer or the user of the BC 19 should

assure that it is used in such an environment.
Immunity test

IEC 60601 test level Compliance level Electromagnetic environment – guidance

Electrostatic discharge (ESD)

IEC 61000-4-2

± 6 kV contact

± 8 kV air

± 6 kV contact

± 8 kV air

Floors should be wood, concrete or ceramic tile. If

floors are covered with synthetic material, the relative

humidity should be at least 30 %.

Power frequency (50 / 60 Hz)

magnetic field IEC 61000-4-8 3 A/m

3 A/m

Power frequency magnetic fields should be at levels

characteristic of a typical location in a typical com-

mercial or hospital environment.

NOTE: U

T

is the a.c. mains voltage prior to application of the test level.