Electromagnetic Compatibility Information

Guidance and manufacture’s declaration – electromagnetic emissions – for all EQUIPMENT and SYSTEMS

Guidance and manufacture’s declaration – electromagnetic emission
The BC80 blood pressure monitor is intended for use in the electromagnetic environment specified below. The customer of the user of the BC80

blood pressure monitor should assure that it is used in such and environment.
Emission test

Compliance Electromagnetic environment – guidance

RF emissions CISPR 11 Group 1

The BC80 blood pressure monitor uses RF energy only for its internal function. Therefore, its RF emissions

are very low and are not likely to cause any interference in nearby electronic equipment.

RF emission CISPR 11

Class B

The BC80 blood pressure monitor is suitable for use in all establishments, including domestic establishments

and those directly connected to the public low-voltage power supply network that supplies buildings used for

domestic purposes.

Guidance and manufacture’s declaration – electromagnetic immunity – for all EQUIPMENT and SYSTEMS

Guidance and manufacture’s declaration – electromagnetic immunity
The BC80 blood pressure monitor is intended for use in the electromagnetic environment specified below. The customer or the user of the BC80

blood pressure monitor should assure that it is used in such an environment.
Immunity test

IEC 60601

test level

Compliance

level

Electromagnetic environment – guidance

Electrostatic discharge (ESD)

IEC 61000-4-2

± 6 kV contact

± 8 kV air

± 6 kV contact

± 8 kV air

Floors should be wood, concrete or ceramic tile. If floor are covered with synthetic

material, the relative humidity should be at least 30%.

Power frequency (50Hz)

magnetic field

IEC 61000-4-8

3 A/m

3 A/m

Power frequency magnetic fields should be at levels characteristic of a typical loca-

tion in a typical commercial or hospital environment.