Beurer BM 44 User Manual

Page 95

Advertising
background image

95

Table 3

For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING

Guidance and manufacturer’s declaration – electromagnetic immunity

The BM 44 is intended for use in the electromagnetic environment specified below. The customer or the user of the BM 44 should assure that it is

used in such an environment.
IMMUNITY

test

IEC 60601

test level

Compliance

level

Electromagnetic environment – guidance

Conducted RF

IEC 61000-4-6

Radiated RF

IEC 61000-4-3

3 V

rms

150 kHz to

80 MHz

3 V/m

80 MHz to

2,5 GHz

3 V

rms

3 V/m

Portable and mobile RF communications equipment should be used no closer to any part of the BM

44, including cables, than the recommended separation distance calculated from the equation appli-

cable to the frequency of the transmitter.
Recommended separation distance:
d = 1,2

d = 1,2

80 MHz to 800 MHz

d = 2,3

800 MHz to 2,5 GHz

Where P is the maximum output power rating of the transmitter in watts (W) according to the trans-

mitter manufacturer and d is the recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey,

a

should

be less than the compliance level in each frequency range.

b

Interference may occur in the vicinity of equipment marked with the following symbol:

NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.

NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures,

objects and people.

a

Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM

and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to

fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the BM 44 is

used exceeds the applicable RF compliance level above, the BM 44 should be observed to verify normal operation. If abnormal performance is

observed, additional measures may be necessary, such as re-orienting or relocating the BM 44.

b

Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.