Guidance and declaration of manufacturer – electromagnetic immunity

Immunity test

IEC 60601 test level

Compliance level

Electromagnetic environment – guidance

Power frequency (50/60 Hz)
magnetic fi eld
IEC 61000-4-8

3 A/m

3 A/m

Power frequency magnetic fi elds should be at levels
characteristic of a typical location in a typical com-
mercial or hospital environment.

NOTE: U

T

is the a.c. mains voltage prior to application of the test level.

Table 3

Guidance and declaration of manufacturer – electromagnetic immunity
Immunity test IEC 60601

test level

Compliance
level

Electromagnetic environment – guidance

Conducted RF
IEC 61000-4-6

Radiated RF
IEC 61000-4-3

3 V

rms

150 kHz to
80 MHz

3 V/m
80 MHz to
2.5 GHz

3V

3 V/m

Portable and mobile RF communications equipment should be used no closer to any part of the
device, including cables, than the recommended separation distance calculated from the equat-
ion applicable to the frequency of the transmitter.

d = 1.2

80 MHz to 800 MHz

d = 2.3

800 MHz to 2.5 GHz

where P is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer and d is the recommended separation distance in meters (m).

Field strengths from fi xed RF transmitters, as determined by an electromagnetic site survey,

a

should be less than the compliance level in each frequency range.

Interference may occur in the vicinity of equipment marked with the following symbol: