Rectifying faults, Specifications – Beurer BM 45 User Manual

Page 26

Advertising
background image

26

• Do not use detergents or solvents.

• On no account must you immerse the computer in water,

otherwise liquid can enter it and cause demage.

9. Rectifying faults

In case of faults, the _ message appears in the display.

Error messages may appear if:

• systolic or diastolic pressure could not be measured (

or

appears on the display),

• systolic or diastolic pressure was outside the measurement

range ( or

Lo appears on the display),

• the cuff is fastened too tightly or loosely (

or

appears

on the display),

• the pump pressure is higher than 300 mmHg (

appears

on the display),

• pumping up takes longer than 160 seconds (

appears on

the display),

• there is a system or device error (

,

,

or

ap-

pears on the display),

• the batteries are almost empty

,

• the data could not be sent to the PC (

appears in the

display).

In such cases, repeat the measurement. Ensure that the cuff

line is correctly attached and that you do not move or speak. If

necessary, reinsert or replace the batteries.

10. Specifications

Model no.

BM 45

Type

M1502

Measurement

method

Oscillometric, non-invasive blood pres-

sure measurement on the upper arm

Measurement range

Cuff pressure 0 – 300 mmHg,

systolic 60 – 260 mmHg,

diastolic 40 –199 mmHg,

Pulse 30 –180 beats/minute

Display accuracy

Systolic ± 3 mmHg, diastolic ± 3 mmHg,

pulse ± 5 % of the value shown

Measurement

inaccuracy

Max. permissible standard deviation

according to clinical testing:

systolic 8 mmHg /diastolic 8 mmHg

Memory

2 x 60 memory spaces

Dimensions

L 165 mm x W 107 mm x H 50 mm

Weight

Approx. 284 g (without batteries)

Cuff size

22 to 36 cm

Permissible operating

conditions

+10 °C to + 40 °C, ≤ 90 % relative air

humidity (non-condensing)

Permissible storage

and transport condi-

tions

- 20 °C to + 55 °C, ≤ 90 % relative air hu-

midity, 800 –1050 hPa ambient pressure

Power supply

4 x 1,5 V

AA batteries

Battery life

For approx. 250 measurements, de-

pending on the blood pressure level

and/or pump pressure