Table 2

For all ME EQUIPMENT and ME SYSTEMS

Guidance and manufacturer’s declaration – electromagnetic immunity
The BM 65 is intended for use in the electromagnetic environment specified below. The customer or the user of the BM 65 should assure that it

is used in such an environment.
IMMUNITY test

IEC 60601 test level

Compliance level

Electromagnetic environment – guidance

Electrostatic discharge (ESD)

IEC 61000-4-2

± 6 kV contact

± 8 kV air

± 6 kV contact

± 8 kV air

Floors should be wood, concrete or ceramic

tile. If floors are covered with synthetic mate-

rial, the relative humidity should be at least

30 %.

Electrical fast transient/burst

IEC 61000-4-4

± 2 kV for power supply lines

± 1 kV for input/output lines

± 2 kV for power supply lines

± 1 kV for input/output lines

Mains power quality should be that of a typical

commercial or hospital environment.

Surge

IEC 61000-4-5

± 1 kV line(s) to line(s)

± 2 kV line(s) to earth

± 1 kV line(s) to line(s)

± 2 kV line(s) to earth

Mains power quality should be that of a typical

commercial or hospital environment.

Voltage dips, short interrup-

tions and voltage variations on

power supply input lines

IEC 61000-4-11

<5 % U

T

(>95 % dip in U

T

)

for 0.5 cycle

40 % U

T

(60 % dip in U

T

)

for 5 cycles

70 % U

T

(30 % dip in U

T

)

for 25 cycles

<5 % U

T

(>95 % dip in U

T

)

for 5 s

<5 % U

T

(>95 % dip in U

T

)

for 0.5 cycle

40 % U

T

(60 % dip in U

T

)

for 5 cycles

70 % U

T

(30 % dip in U

T

)

for 25 cycles

<5 % U

T

(>95 % dip in U

T

)

for 5 s

Mains power quality should be that of a typi-

cal commercial or hospital environment. If

the user of the BM 65 requires continued

operation during power mains interruptions,

it is recommended that the BM 65 be powe-

red from an uninterruptible power supply or

a battery.

Power frequency (50/60 Hz)

magnetic field IEC 61000-4-8

3 A/m

3 A/m

Power frequency magnetic fields should be at

levels characteristic of a typical location in a

typical commercial or hospital environment.

NOTE: U

T

is the a.c. mains voltage prior to application of the test level.