Technical alarm – description

Should the recorded blood pressure (systolic or diastolic)

lie outside the limits specified in the section “Technical

specifications”, the technical alarm will appear on the dis-

play indicating either “ ” or “ ”. In such cases, you should

seek medical assistance and check the accuracy of your

procedure.

The limit values for the technical alarm are factory set and

cannot be adjusted or deactivated. These alarm limit values

are accorded second priority under the standard IEC 60601-

1-8.

The technical alarm is a non-locking alarm and must not be

reset. The signal shown on the display will disappear auto-

matically after about 8 seconds.

9. Cleaning and storing the instrument

• Clean your blood pressure computer carefully only with a

slightly moistened cloth.

• Do not use detergents or solvents.

• On no account must you immerse the computer in water,

otherwise liquid can enter it and cause demage.

• When storing the device, make sure that no heavy objects

are placed on top of it. Remove the batteries. The cuff

tube should not have any sharp kinks.

10. Specifications

Model no.

BM 65

Measurement

method

Oscillometric, non-invasive blood pres-

sure measurement on the upper arm

Measurement

range

Cuff pressure 0 – 300 mmHg,

systolic 60 – 260 mmHg,

diastolic 40 –199 mmHg,

Pulse 40 –180 beats/minute

Display accuracy

Systolic ± 3 mmHg,

diastolic ± 3 mmHg,

pulse ± 5 % of the value shown

Measurement inac-

curacy

Max. permissible standard deviation

according to clinical testing:

systolic 8 mmHg /diastolic 8 mmHg

Memory

10 x 30 memory spaces

Dimensions

L 240 mm x W 120 mm x H 50 mm

Weight

Approx. 419 g (without batteries)

Cuff size

22 – 30 cm / 30 – 42 cm

Permissible operat-

ing conditions

+10 °C to +40 °C, ≤ 90 % relative air

humidity (non-condensing)

Permissible storage

conditions

-20 °C to +55 °C, ≤ 90 % relative air

humidity, 800 –1050 hPa ambient

pressure