Flowserve CPXM User Manual

Page 20

Advertising
background image

CPXM and CPXRM USER INSTRUCTIONS ENGLISH 71569101 10-08

Page 20 of 32

flowserve.com

g) Carefully loosen and back off the grubscrew and

tighten the coupling bolts, ensuring that the gap
is equal between the coupling halves.

h) Torque the screws to the specified values:

M 8 - 30 Nm (22 lbf•ft)
M 10 - 58 Nm (43 lbf•ft)

i)

If a cartridge seal is fitted it should be reset at this
point.

6.8 Disassembly

Refer to Safety section before dismantling the

pump.

Before dismantling the pump for

overhaul, ensure genuine Flowserve replacement
parts are available.

Refer to sectional drawings for part numbers and
identification. See section 8, Parts lists and drawings.

6.8.1 General
a) Close suction and discharge valves and drain

liquid from the pump.

b) Remove screws from pump casing and pull motor

complete with rotating assembly from back of
pump casing, which will be left connected in
position in the pipework.

c) Unclip the coupling guards.
d) Take out the muff coupling screws and remove

coupling.


6.8.2 Pumps with single seals
a) Prevent the motor shaft from rotating.
b) Carefully rotate the impeller in an anti-clockwise

rotation, whilst supporting the impeller, until it
releases from the adjustment stud. Take care
not to damage the seal.

c) Withdraw the impeller/stubshaft assembly,

complete with mechanical seal, from the seal
housing.

d) Release the tension in the mechanical seal. The

seal manufacturer's instructions should be
followed for dismantling and assembling the seal.

e) Remove seal housing bolts and withdraw the seal

housing from the bracket.


6.8.3 Pumps with double mechanical seals
a) Remove the nuts retaining the seal housing to

the bracket.

b) Carefully rotate the impeller in an anti-clockwise

direction whilst supporting the impeller and seal
housing. The integral impeller and stub shaft
combined with the seal housing will release from
the unit as a sub-assembly. (Large pump sizes
have a tapped hole for fitting a lifting eye to assist
with this procedure.)

c) Remove the nuts from the seal cover to gain

access to the seals. Release the tension in the
mechanical seals.

d) The seal manufacturer's instructions should be

followed for dismantling and assembling the seal.

e) The mounting bracket can now be removed from

the motor after first removing the fixing screws.

f) The adjustment stud in the motor shaft can be

removed using two M 8 nuts.

6.9 Examination of parts

Used parts must be inspected before

assembly to ensure the pump will subsequently run
properly. In particular, fault diagnosis is essential to
enhance pump and plant reliability.

6.9.1 Casing, seal housing and integral impeller/
stubshaft assembly
Inspect for excessive wear, pitting, corrosion, erosion
or damage and any sealing surface irregularities.
Replace as necessary.

6.9.2 Integral impeller/stubshaft
Replace if the shaft is grooved or pitted or if the
impeller vanes are eroded or damaged.

6.9.3 Gaskets
After dismantling, discard and replace. (If the pump
seal arrangement has a lip seal, it should be replaced
at overhaul.)

6.9.4 Motor
Check the motor shaft for free rotation and absence
of bearing noise or shaft 'float'. If necessary, have
the motor serviced by a specialist or replace with a
Flowserve approved type.

Advertising
This manual is related to the following products: