Flowserve ISC2 Single Pusher Repair User Manual

Page 7

Advertising
background image

7

6.17 Perform a dry fit of the collar on the sleeve.
6.18 Install the drive collar [58] onto the sleeve

assembly [1]. The drive collar may need to be

rotated so that the set screws [57] line up with the

large holes and quarter-dog set screws [57.1] line

up with the two smaller holes.

Note: Some sizes > 2.750" do not utilize quarter-

dog set screws.

6.19 Install the setting devices and cap screws [103]

and [40] into the drive collar, engaging with the

gland [11]. See Figure 17.

Caution: Do not rotate the gland assembly

during installation. This may cause the springs

[16] installed in step 6.12 to buckle if the gland

assembly [11] is rotated with respect to the

stationary face support [100].

6.20 While compressing seal by pushing down on

gland assembly [11], tighten the quarter-dog set

screws [57.1] to engage into the two smaller holes

of the sleeve assembly [1]. If the seal does not

contain quarter-dog set screws then install the set

screws [57] into the sleeve [1]. See Figure 18.

Caution: Over tightening will cause distortion

of the sleeve assembly [1]. Check integrity of

the sleeve with a plug of the appropriate size to

ensure no distortion has occurred.

6.21 Install sleeve O-ring [19] into the inner diameter

groove of the sleeve assembly.

6.22 The cartridge seal assembly is now ready for

testing.

6.23 Adhere the gland gasket [18] to the gland gasket

surface with a spray adhesive such as 3M Super

77

®

. See Figure 19.

6.24 Permanently mark the seal type ISC2-PX or

ISC2-XP, seal size and gland ring

material clearly on the gland surface.

See Figure 1 for placement location.

Figure 16

Figure 17

Figure 18

ISC2-PX

M060

316 SS

Example

inch

marking

ISC2-PX

1.875

316 SS

Example

metric

marking

Figure 19

Advertising