Naccurate or, Issing, Ecorded on – Flowserve 103 Smart Multi-Port IPS Wireless User Manual

Page 15: Vb-103-tam, Ights on, Vb-103-vis

Advertising
background image

IPS WIRELESS SELD-103 ENGLISH 26999976 11-13

Page 15 of 20

flowserve.com

If data is being read from this location, then ensure
that the repeater network is appropriately located
and functional.

9.1.4 Receiver Not Configured Correctly
Check receiver configuration to ensure it is set up
to receive data from this particular SELD-103 unit
(refer to receiver user instructions).

If none of the above solutions are successful in
getting data reception at the receiver, contact the
factory for additional support.

9.2 Inaccurate or Missing Data

If the data being read from the SELD-103 unit is
either inaccurate or missing, this could have
several causes:

Loose cable connections

Unit configuration incorrect

Inaccurate sensor calibration

Incorrect receiver configuration

9.2.1 Loose Cable Connections
Ensure the sensor cables to and sensors ports on
the SELD-103 unit are free from dust and debris.
Use compressed moisture free air to blow out any
debris if needed.

Reconnect all sensor cables to the SELD-103 unit
to ensure tight connections.

9.2.2 Unit Configuration Incorrect
If the inaccurate/missing data problem is across all
sensors, reload the original configuration for the
SELD-103 unit (see PSP user instructions).

9.2.3 Inaccurate Sensor Calibration
If the data received on certain channels/sensors is
not accurate, recalibrate those sensors (see PSP
user instructions).

9.2.4 Incorrect Receiver Configuration
Refer to receiver instructions to confirm correct
configuration. If necessary, contact factory to
reload receiver configuration settings.

9.3 No Data Recorded on VB-103-TAM

If no data is being stored to the VB-103-TAM unit,
this could have several causes:

Loose sensor cable connections

Loose VB-103-TAM connection

Low battery on VB-103-TAM

Incorrect VB-103-TAM log interval
configuration

9.3.1 Loose Sensor Cable Connections
Ensure the sensor cables to and sensors ports on
the SELD-103 unit are free from dust and debris.
Use compressed moisture free air to blow out any
debris if needed.

Reconnect all sensor cables to the SELD-103 unit
to ensure tight connections.

9.3.2 Loose VB-103-TAM Connection
Ensure the VB-103-TAM and SELD-103 ports are
free from dust and debris. Use compressed
moisture free air to blow out any debris if needed.

Disconnect and reconnect the VB-103-TAM from
the SELD-103 unit. Ensure that VB-103-TAM is
completely and fully threaded onto the SELD-103.

9.3.3 Low Battery on VB-103-TAM
Check to see if VB-103-TAM unit can be read on
an IPS Dock. If not, replace VB-103-TAM unit (the
VB-103-TAM battery is not user replaceable).

9.3.4 Incorrect VB-103-TAM Log Interval
Configuration
Connect the VB-103-TAM to the IPS Dock and
ensure the log interval is set to the expected value
(see IPS Dock user instructions for additional
details on how to perform this task).

9.4 No Lights on VB-103-VIS

If no lights are flashing on the VB-103-VIS, this
could have several causes:

Loose VB-103-VIS connection

9.4.1 Loose VB-103-VIS Connection
Ensure the VB-103-VIS and SELD-103 ports are
free from dust and debris. Use compressed
moisture free air to blow out any debris if needed.

Disconnect and reconnect the VB-103-VIS from the
SELD-103 unit. Ensure that VB-103-VIS is
completely and fully threaded onto the 103.

If none of the above solutions are successful in
resolving the issue, contact the factory for
additional support.