Fluke Biomedical 6000D Impulse Getting Started User Manual

Page 31

Advertising
background image

Defibrillator/Transcutaneous Pacer Analyzer

Specifications

19

Noise Immunity

Wave ................................................................... Sine
Line Frequency....................................................50 or 60 Hz (± 0.5 Hz)
Amplitude ............................................................0.0 to 10.0 (by 0.5) mV
Accuracy..............................................................± 5%

Transvenous Pacer Pulse Simulation

Widths

Range..............................................................0.1, 0.2, 0.5, 1.0, and 2.0 ms
Accuracy..........................................................±5 % of setting

Amplitudes...........................................................0 (off) and ±2, ±4, ±6, ±8, ±10, ±12, ±14, ±16, ±18, ±20, ±50, ±100, ±200, ±500, and

±700 mV

Accuracy..............................................................± (10 % of setting + 0.2 mV)

Arrhythmia Selections

Pacer Interactive (Transcutaneous pacer, Impulse 7000DP only)

Demand...........................................................30 to 360 (by 1) BPM
Asynchronous
Non-Capture
Non-Function
Threshold (Interactive pacing simulation only) 10 to 250 (by 10) mA

Supraventricular

Atrial Fibrillation Coarse
Atrial Fibrillation fine
Atrial Flutter
Sinus Arrhythmia
Missed Beat
Atrial Tachycardia

Advertising
This manual is related to the following products: