Test load selections, Measurements – Fluke Biomedical 6000D Impulse Getting Started User Manual

Page 34

Advertising
background image

Impulse 6000D, 7000DP
Getting Started Manual

22

Transcutaneous Pacemaker Analyzer Specifications (Impulse 7000DP only)

Test Load Selections

Defibrillator Input

Fixed Load ..........................................................50 Ω
Accuracy .............................................................

±1 %, non-inductive (<2 μH)

Power Rating.......................................................10 defib pulses of 360 J every 5 minutes

Pacemaker Input

Variable Load ......................................................50 to 1500

Ω in 50 Ω steps

Accuracy .............................................................

±1 %, non-inductive (<2 μH)

Power Rating.......................................................5 W (average), 40 W (peak) @ 1000 Ω

Measurements

Manufacturer Specific Algorithms

GE Responder (1500 & 1700)
MDE 300 (Medical Data Electronics)
Medtronic ERS/Physio Control LIFEPAK
MRL (Medical Research Laboratory/Welch Allyn)
Philips/Agilent/HP
Schiller Medical
ZOLL Medical
(plus a general purpose default algorithm selection)

Current

Range..................................................................4.00 to 250 mA
Accuracy .............................................................±(1% of reading + 0.02 mA)

Advertising
This manual is related to the following products: