3.3 Replacing the Mylar Cover

If the step tablet mylar cover is damaged, replace it using the following procedure:

1. Remove the batteries (refer to Replacing the Batteries, Section 3.1).

2. Remove the case from the base by removing the four (4) screws on the sides of the case.

3. Remove the cover and hinge from the case by removing the two (2) screws located inside the case.

4. Starting at the edge nearest the cover, carefully peel off the mylar film.


Be careful not to lift off the step tablet or the
double-sided tape.


5. Clean the step tablet with a camel hairbrush.

6. Position the replacement the mylar cover film (Part No. 138032) over the step tablet.

7. Secure the sides and front edge of the film to the step wedge using Magic

TM

transparent tape, 0.5

inches wide.