Indicators and controls, 1 front panel – Fluke Biomedical 07-479 User Manual

Page 13

Advertising
background image

Indicators and Controls

Front Panel

4

4-1

Section 4

Indicators and Controls

4.1 Front Panel

See Figure 4-1 for front panel control and indicator location.

Numeric Readout

The front panel 3-½ numeric digit LCD readout displays the measurement data. Also indicated on the
display are low battery, overrange kVp, underrange kVp, high mAs, and low mAs conditions as listed in

Table 4-1.

Figure 4-1. Front Panel Indicators and Controls

Table 4-1. Displayed Indications

Condition Indication

Low Battery

LO BAT in the upper left corner of the display

Overrange kVp

Flashing decimal point directly above A

Underrange kVp

Flashing decimal point directly above B.

High mAs

Flashing decimal point directly above C/D and a non zero numeric display

Low mAs

Flashing decimal point directly above C/D and a display numerically equal
to zero


An overrange kVp condition indicates that the measured kVp is above the calibrated range. An
underrange kVp condition indicates that the measured kVp is below the calibrated range.

Advertising