Introduction, 1 product description, 2 specifications – Fluke Biomedical 07-479 User Manual

Page 5

Advertising
background image

Introduction

Product Description

1

1-1

Section 1

Introduction

1.1 Product Description

The Model 07-479 Dental kVp Meter is designed to measure the effective peak potential applied to dental
type x-ray tubes. The measurement is made without direct connection to the x-ray unit being tested. The
kVp Meter is a battery powered, self-contained instrument designed for portable use.

The kVp meter uses two differentially filtered detectors whose ratio of integrated outputs is calibrated over
the range of 45 to 90 kVp. Measurement results are presented on an easy to read 3 ½ digit liquid Crystal
Display. Status is indicated by decimal point positioning on the readout. Low battery condition is indicated
by LO BAT displayed on the readout.

1.2 Specifications

Range

45 to 90 kVp

Accuracy

± 3 kV or 3%, whichever is greater.

Calibrated with NBS traceable voltage divider.

Reproducibility

± 0.5 kV

Resolution 0.1

kV

Distance Dependence

Negligible for SSD of 20 cm or greater

Angle Dependence

Negligible for 10 degrees or less

Position Dependence

Negligible for misalignment of 5 cm or less at 30 cm

Calibration Period One

year

Minimum Requirements

5 mAs @ 45 kVp (20 cm)

0.03 mAs @ 90 kVp (20 cm)

Power Requirements

9 V Alkaline Battery, MN1604A or equivalent

Display

3 ½ Digit LCD

Low Battery Indication: LO BAT

High kV Indication: Left blinking decimal

Lo kV Indication: Middle blinking decimal

High Intensity Indication: Right blinking decimal and non-zero display

Low mAs Indication: Right blinking decimal and zero display

LED Indicator

Auto Reset, new exposure detected, and last reading cleared.

Controls

Front Panel: Power switch and phase selection switch.

Output Connector

BNC Connector on rear panel for waveform output.