Calibration and adjustments, 1 lo battery threshold adjustment, 2 kvp calibration adjustments – Fluke Biomedical 07-479 User Manual

Page 15

Advertising
background image

Calibration and Adjustments

Lo Battery Threshold Adjustment

5

5-1

Section 5

Calibration and Adjustments

The Model 07-479 kVp Meter has been factory calibrated. Any calibration should be done by the factory
or a qualified calibration facility. The unit should be calibrated once a year by adjusting the five variable
resistors as discussed in the following sections.

5.1 Lo Battery Threshold Adjustment

Resistor R25 adjusts the threshold for the LO BAT indication on the LCD Display. Use the following
procedure to adjust R25:

1. Remove

the

battery.

2. Connect a variable DC power supply to the instrument at the battery connector cable.

3. Connect a voltmeter between battery plus (TP3) and analog ground (TP4).

4. Turn on the power supply and adjust it to 9 V.

5. Turn on the kVp meter. The voltmeter should read between 2.4 and 3.2 V.

6. Slowly decrease the power supply until the voltage across the test points just returns to the value it

had prior to dropping off.

7. Turn R25 until LO BAT disappears from the display.

8. Turn R25 slowly in the opposite direction just until LO BAT reappears on the display.

9. Connect the voltmeter between analog ground and the power supply negative terminal.

10. Increase the power supply voltage until the voltmeter reads 4.5 V while being careful not to exceed

9 V from the power supply. LO BAT should be off at this point.

5.2 kVp Calibration Adjustments

The following are some guidelines to ensure proper calibration of the kVp meter:

1. An independent accurate means of measuring the x-ray machines is necessary.

2. The kVp meter must be calibrated at 50 kVp and 90 kVp (± 1 kVp) if it is to be within specification

over its entire range.

3. The phase switch must be in correct position during each portion of the calibration.

4. The potentiometers must be adjusted interactively as detailed below.

5. Be sure to follow the procedure outlined below in the order listed.


Use the following procedure for kVp calibration:

1. Set a calibrated dental x-ray machine to 50 kVp with sufficient mAs to obtain a reading.

2. Locate the Model 07-479 kVp meter in the center of the beam

with the cover removed from the

meter.