Victoreen 07-479
Operators Manual

5-2


Be sure the phase switch on the front panel of the
kVp meter is in the appropriate position. Adjust R27
fully counter clockwise.

3. Make an exposure.

4. Adjust R6 (offset adjustment) until the kVp meter reads 50 kVp.

5. Set the x-ray machine to 90 kVp.

6. Make an exposure.

7. Adjust R5 (slope adjustment) until the kVp reads 90 kVp.

8. Repeat Steps 1 through 7 until the readings are accurate.

NOTE