6 RECM Alarm Test Element

6.1 Introduction

This test ensures that the ESU will sound an alarm if the resistance between the two electrodes exceeds a
specified limit. The QA-ES will gradually increase the resistance, starting at 10 Ohm and increasing through all
available load settings. At a certain value, the ESU should sound an alarm. It is this value that will be returned.

The user must define the acceptable alarm level through the expected results dialog. On this comparison the test
will either pass or fail.

6.2 Custom

setup

6.2.1 Software

Settings

The software settings are as section 5.2.

6.2.2 Hardware

Settings

The hardware settings are found on the left of the custom setup dialog within the test settings box.

Delay

This is the time between the loads being applied to the ESU.

6.3 Expected

Results

This performance test will contain the acceptable upper limit for the alarm to trigger.