4 output test element, 1 introduction, 2 custom setup – Fluke Biomedical Ansur QA-ES plug-in Version 1.4.4 User Manual

Page 8

Advertising
background image

8

4 Output Test Element

4.1 Introduction

Within this test the user has the opportunity to determine the power performance of an ESU over a desired range
of power values with a constant load across its electrodes.

The user must define a desired power value for every power measurement to be made. This is achieved through
the custom setup dialog.

The user must also define the acceptable deviation to the desired power performance, facilitated on the expected
results
dialog. If one measurement fails to meet these limits then the test will fail.

4.2 Custom

setup

4.2.1 Software

Settings

The software settings are displayed in the left of the custom setup dialog.

Enable Skip Button

If set to False (check box clicked so that no checkmark is displayed) the Skip button in
the Test Guide (see section 7.1) for this test will be disabled.

Enable NA Button

If set to False (check box clicked so that no checkmark is displayed) the NA button in the
Test Guide (see section 7.1) for this test will be disabled.

Report Crest Factor The measured crest factor to be shown in the final report.

Report Current

The measured current to be shown in the final report.

Report Peak-to-
peak voltage

The measured Peak-to-peak voltage

to be shown in the final report.

4.2.2 Hardware

Settings

The hardware settings are defined in two distinct parts. The first of these is found on the left of the custom setup
dialog within the test settings box.