3 expected results – Fluke Biomedical Ansur QA-ES plug-in Version 1.4.4 User Manual

Page 9

Advertising
background image

9

Delay

This is the delay between the changing of the loads and taking a power measurement.

Mode

Defines the mode that the ESU is to be in. The mode determines the type of waveform applied
to the electrodes.

Load Setting

This is the constant load value for when taking every power output measurement. User can
enter the load values or can select a valid load value from the pop up help list box which
appears next to the edit box.


The second part is found on the right of the custom setup dialog within the test details box.

This determines the number and range of power measurements to be taken. Each line defines a desired power
value for a single performance measurement. The toolbar allows addition and deletion of individual settings. If
the

symbol appears then that particular measurement setting is invalid.


Expected
Power

This is the power value on the ESU required for the test. Each load setting will have a unique
power value. Naturally this is the expected power measurement.

ESU Power
Setting

An optional entry that gives the dial setting appropriate to the power value.

4.3 Expected

Results

Every measurement returns a power, current, crest factor and voltage peak-to-peak value and so limits can be
placed against any of them. A drop down list enables the user to choose the value for which the limit applies (see
below). Limits can be added or deleted against a standard by right clicking the mouse control when the pointer is
over a limit. The returned value must always be within the upper and lower limit, where stipulated, for the
complete test.

The exception is the power value. The user defined power value in the test details box against each
measurement is the desired value, naturally. The limits define the acceptable range around this desired power
value.

Advertising