QA-ES Automation Bundle
Users Manual

10

HF Leakage Current Tests

These tests measure the leakage current from the ESU under test. Figures 5 through 10 show the connections used to make
the different HF leakage current measurements on the unit under test.

ESU

QA-ES II

Active

Coag (Black)

Cut (Yellow)

F

2

0

0

Neutral/Disp.

Lead

3 or 4

Lead 1

Lead 5

giw02.eps

Figure 5. HF Leakage Current on Active Electrode for HF Insulated ESUs (CF Type) Connection - Monopolar