Qa-es automation bundle users manual – Fluke Biomedical QA-ES Automation Bundle User Manual

Page 20

Advertising
background image

QA-ES Automation Bundle
Users Manual

12

Active

Coag (Black)

Cut (Yellow)

Neutral/Disp.

2

0

0

ESU

QA-ES II

Lead

3 or 4

Lead 1

Lead 5

giw04.eps

Figure 7. HF Leakage Current (Fig 101) for HF Grounded ESUs (BF Type) Connections - Monopolar