Fetal Simulator

Specifications

7

Uterine Activity

Note

TOCO waveform selection is not available during TREND #1.

Table 2. Uterine Activity

Activity Description

EXECUTE WAVEFORM Start

TOCO

waveform

UTERINE WAVE OFF Stop

TOCO

waveform

ANALOG 0 TO +1 VOLT

Analog range 0 to +1 V (1 V=100 mmHg)

ANALOG 0 TO -1 VOLT

Analog range 0 to -1 V (-1 V=100 mmHg)

UTERINE WAVE 0-25

Range of TOCO waveform

UTERINE WAVE 0-50

Range of TOCO waveform

UTERINE WAVE 0-100

Range of TOCO waveform

SHORT DURATION

TOCO waveform of short duration

NORMAL DURATION

Normal duration of TOCO waveform

INCREASED DURATION

Long duration of TOCO waveform

UTERINE LEVEL =ZERO

Zero TOCO channel (automatic on power up)

UTERINE STATIC +20

Increase TOCO static level by 20 mmHg (0-100 mmHg)

INCR. RESTING TONE

Resting tone increases

COUPLING

2 close TOCO waves

TRIPLING

3 close TOCO waves

UTERINE PRESSURE SENSIVITY

5 or 40 μV on power up