Fluke Biomedical PS320 User Manual
Page 23

Fetal Simulator
Simulating Functions
15
Table 4. Simulating Functions (cont.)
Main Menu
Sub Menu
Function
*MAT SEN. 0.5mV
Sets maternal ECG amplitude to 0.5 mV
*MAT SEN. 1 mV
Sets maternal ECG amplitude to 1 mV
*MAT SEN. 2 mV
Sets maternal ECG amplitude to 2 mV
*MAT RATE +20
Increases maternal rate by 20 bpm
→MATERNAL ECG
*MAT RATE –20
Decreases maternal rate by 20 bpm
*SENS @ 5
μV mmHg
Sets pressure sensitivity to 5
μV
*SENS @40
μV mmHg
Sets pressure sensitivity to 40
μV
*UTERINE LVL =0
Sets uterine level to 0 (zero)
*ANALOG TO +1 V
Sets analog uterine output from 0 to +1V. 1V=100 mmHg
*ANALOG TO –1V
Sets analog uterine output from 0 to –1V. -1V = 100 mmHg
*UTERINE STATIC
Sets uterine output in steps of 20 mmHg
*EXECUTE TOCO WA
Starts a TOCO wave, not valid during trend selection
*UTERINE WVF OFF
Stops TOCO wave, not valid during trend selection
→UTERINE ACTIVITY
*UTERINE WV 0-25
Sets standard TOCO wave from 0 to 25 units