2. specifications – Fluke Biomedical 07-440 User Manual

Page 6

Advertising
background image

Nuclear Associates 07-440
Operators Manual

1-2

1-2. Specifications

Table 1-1. Instrument Specifications

Feature

Specifications

Density Range

0-4.50

Response Time

<2 seconds at optical density of 3.00

Warm-Up Time

10 seconds maximum

Accuracy

± 0.02 at density of 3.00 with 1 mm aperture

Density Subtraction

Zero control permits density subtraction of up to 4.50 with 3 mm aperture. Minus sign
included on display to show negative relative densities.

Zero Control

10 turn control permits exact zeroing with flashing minus sign indicating correct zero
to within 0.002.

Zero Drift

Negligible

Spectral Response

Centered at approximately 465 nanometers

Supplied Apertures

3 mm, 2 mm, and 1 mm diameter Aluminum

Throat

210 mm (8.25 in.)

Illuminated Table Area

120 x 140 mm (4.75 x 5.5 in.)

Detector

Silicon diode

Detector Light Source

White LED provides extremely long life with minimum spectral and intensity
degradation. Also reduce specimen heating to a minimum.

Readout

Located on measurement arm; three L.E.D. digits with minus sign, 10 mm (0.4 in.)
high.

Dimensions
(H x W x D)

4.9 in. x 6.3 in. x 13.4 in.
(125 mm x 160 mm x 340 mm)

Weight

7.5 lbs. (3.4 kg)

Shipping Weight

10 lbs. (4.3 kg)

Power Switch

Selectable

115 VAC, 50/60 HZ, 15 W
230 VAC Unit available
Order #07-440-20
Fuses
For 115 VAC - Use 1.0 A, 250 V, AGC Fast Acting
For 230 VAC - Use 0.5 A, 250 V, AGC Fast Acting

Operating Conditions

10° C to 40° C (50° F to 104° F)
Maximum 90% relative humidity, non-condensing

RS-232C Interface

User selectable baud rates ASCII format

Supplied Accessories

Part No. 010128 Step Tablet for calibration
Four #37 lamps
One extra O-ring light seal
One set of three apertures