Fluke Biomedical 07-440 User Manual

Page 9

Advertising
background image

Operation

Setup

2

2-3

ZERO SET

1. Place the supplied 3 mm aperture in position under the photohead, in the depression over the filter

light source WITH THE FLAT SIDE UP.

2. Position the ZERO Control fully counterclockwise.

3. Apply pressure to the measurement arm as indicated (PRESS HERE) until it makes firm contact

with the aperture. The display should read 0.010 to 0.20 (positive).

4. Turn the ZERO control clockwise until the minus sign (located just above the left hand digit) flashes

on and off, indicating an exact zero condition.

BAUD RATE SELECTION

The unit is shipped with the baud rate factory set for 9600, 8 data bits with one stop bit. Use the following
procedure to change the baud rate and/or data bits, stop bit, and parity combination.

1. Remove the four (4) screws on white panel containing the power switch.

2. Carefully slide the panel forward to gain access to the dip switches.

3. Position SW1-3 to SW1-6 as indicated in Table 2-1 for the desired baud rate selection.

4. Position SW1-1 and SW1-2 as indicated in Table 2-2 for the desired data bit, stop bit, and parity

combination.

5. RS-232 Cable Connections are as follows:

a. Pin 2: Transmit

b. Pin 3: Receive (Not Used)

b. Pin 7: Signal Ground

Table 2-1. Baud Rate Selection

Switch SW1 - Position

Baud Rate

6 5 4 3

300 OFF

OFF

ON

OFF

1200 OFF ON OFF OFF

2400

ON OFF OFF OFF

4800 OFF ON ON OFF

9600 OFF ON ON ON

Table 2-2. Data Bit, Stop Bits, and Parity Selection

Switch SW1 - Position

Selection

1 2

8 DATA, 2 STOP, NO PARITY

OFF

OFF

8 DATA, 1 STOP, NO PARITY

OFF

ON

7 DATA, 2 STOP, NO PARITY

ON

OFF

7 DATA, 1 STOP, NO PARITY

ON

ON