Victoreen 1060AM
Operators Manual

3-2

3.2.1 COMMAND SET

The COMMAND SET is comprised of STANDARD, SETUP, TEST, and RADIATION commands.

STANDARD commands are always available to all communication ports.

SETUP commands (marked "s") are used to configure the 1060 and are made available through switch
S2 as follows:

Port 1 (R-S232), S2, Position 2 ON = Setup commands available

Port 2 (RS-485), S2, Position 3 ON = Setup commands available

Port 3 (RS-485), S2, Position 4 ON = Setup commands available

SETUP MODE disables RADIATION commands.

SETUP MODE cannot be entered while Pulse Output is active.


TEST commands (marked "t") are used to test parts of the hardware.

At this time, only the hardware that is exercised is the analog output.

TEST MODE and SETUP MODE can both be active.

TEST MODE can be entered while Pulse Output is active.


RADIATION commands (marked "r") are used to poll the 1060 for specific radiation data.

To enter Setup mode, enter the command ETM. This will enable the TEST and disable the RADIATION
commands. The analog output driver is disabled and the user can drive the output via commands ANS,
DAC, and LOG (see the following COMMAND SET list). To exit TEST mode, enter XTM. This will disable
the TEST and enable the RADIATION commands. The analog output driver is enabled.

ANS

t

Analog out per Counts per Second entered

CPH

r

Counts per Hour (read counts averaged over 60 min)

CPM

r

Counts per Minute (read counts averaged over 60 sec)

CPS

r

Counts per Second (read latest second count value)

DAC

t

DAC POT output per numeric entry 0 to 255

ESM

r

Enter Setup Mode (enable Setup Commands, disable Radiation

Commands) Pulse output cannot be active

ETM

t

Enter Test Mode (enable Test Commands and Radiation Commands)

Analog meter is controlled by Test Commands.

LOG

t

Drive analog meter numeric input, i.e. LOG 10, needle moves to position 10.

MHV

Measure High Voltage through Analog to Digital converter)

RxH

r

Counts per Hour (using Conversion Factor x, where x=0 to 9)

RxM

r

Counts per Minute (using Conversion Factor x, where x=0 to 9)

RxS

r

Counts per Second (using Conversion Factor x, where x=0 to 9)

RAB

Reset Averaging Buffers (Resets counting buffers)

RCx

Read Conversion Factor x, where x=0 to 9

RCR

Read Conversion Factor CPM to mR/Hr

RDI

Read Digital Input (Reads 16 digital inputs)

RDO

Read Digital Output (Reads 16 digital outputs)

RGT

Read GM Threshold setting from EEPROM

RHD

Read High Discriminator Voltage setting from EEPROM

RHS

Read High Scale analog output (shown in units of mR/hr.)

RHV

Read High Voltage (Reads HV setting from EEPROM)

RIC

Read Input Control byte (detector input channel)

RID

Read 1060 unit Identification

RLD

Read Low Discriminator Voltage setting from EEPROM

RLS

Read Low Scale analog output (shown in units of mR/hr.)

RRH

r

Read mR/Hr (averaged over 60 minutes)

RRM

r

Read mR/Hr (averaged over 60 seconds)