Operation

Terminal Operation

3

3-3

NOTE

RRS

r

Read mR/Hr (latest second mR/Hr value)

RSH

r

Read Sieverts (averaged over 60 minutes)

RSM

r

Read Sieverts (averaged over 60 seconds)

RSS

r

Read Sieverts (latest second Sievert value)

RTA

Read TAU (deadtime)

SCx

s

Set Conversion Factor x, where x = 0 to 9

*SCR

s

Set Conversion Factor CPM to mR/Hr (enter CPM per mR/hr)

SDO

Set Digital Output (Sets 16 digital outputs)

SFD

s

Set output Format Decimal (integer)

SFE

s

Set output Format Engineering (default, x.xxExx)

SFH

s

Set output Format Hex (integer, 16 hex digits)

SGT

s

Set GM Threshold (Sets Thresh., stores in EEPROM)

SHD

s

Set High Discriminator (Sets Hi Discrim., stores in EEPROM)

SHS

s

Set High Scale analog output (entered in units of mR/hr)

SHV

s

Set High Voltage (Sets HV, stores setting in EEPROM)

SIC

s

Set Input Control

SID

s

Set unit Identification

SLD

s

Set Low Discriminator (Sets Lo Discrim., stores in EEPROM)

SLS

s

Set Low High Scale analog output (enter in units of mR/hr)

*STA

s

Set TAU (deadtime correction);

VER

Read firmware Version and Revision level

XSM

s

Exit Setup Mode

XTM

t

Exit Test Mode



Commands SCR and STA must be used to enter
the CPM to mR/hr conversion factor and deadtime
correction. These values must be entered
whenever a new detector is installed. These values
can be found on the detector calibration sheet.

3.3 LED Definitions

1.

Heartbeat (flash pulse indicates proper operation)

2.

HV ON

3.

RS-232 (Port 1) Command accepted and processed

4.

RS-232 (Port 1) Character received

5.

RS-485 (Port 2) Command accepted and processed

6.

RS-485 (Port 2) Character received

7.

RS-485 (Port 3) Command accepted and processed

8.

RS-485 (Port 3) Character received

3.4 DIP Switch Definitions

1.

ON = HV ON

2.

ON = Port 1 (RS-232) Setup commands available

3.

ON = Port 2 (RS-485) Setup commands available

4.

ON = Port 3 (RS-485) Setup commands available

5.

Open

6.

Open